LC-MS Calculators
Method scaling, column volumes, adducts and mass accuracy, and dilutions. Results update as you type.
Scale a Method to a New Column
Keeps linear velocity and gradient column volumes the same. Same stationary phase assumed.
Current Method
New Column
Scaled Method
Pressure and plate numbers are relative estimates at the same solvent, temperature, and packing quality. Check your system's pressure limit before running.
Column Volume and Wash Volumes
Empty column volume = π × r² × L. Use it to turn "10 column volumes" into mL and minutes.
Column and Flow
Results
Quick Reference at This Flow Rate
| Column volumes | Volume (mL) | Time (min) |
|---|
What the Sources Say
- Between gradient runs, Waters recommends 3 to 5 column volumes to clean a column (counting the system dwell volume) and 5 to 10 empty column volumes of re-equilibration for a reversed-phase column. Waters WKB51034
- ACE advises at least 10 column volumes to equilibrate a reversed-phase column before analysis. For HILIC gradients, about 10 column volumes after a run is usually enough, a new HILIC column needs 60 to 80, and buffered mobile phases can take longer. ACE Knowledge Note 0025
- Ion exchange phases can need up to 100 column volumes, while some silica columns are stable with 3 to 5. MicroSolv suggests starting with extra time during development and shortening it until retention precision begins to suffer. MicroSolv KB
- Empty volume is π × r² × L, and a void-volume estimate multiplies that by about 0.66 for fully porous and 0.49 for superficially porous packing. Measuring an unretained marker gives the true value. Waters WKB48971 · WKB28079
So 10 column volumes is a sensible default for reversed-phase re-equilibration, at the upper end of Waters' 5 to 10, but it is not universal. Sources also count differently: Waters uses empty column volumes, while ACE's table of internal volumes works out to about 63% of the empty-tube volume. Confirm the number for your own method.
Adducts and Mass Error
Adducts from Neutral Monoisotopic Mass
| Ion | m/z |
|---|
Uses ion masses (electron mass accounted for). Verify against your own software before reporting values.
Mass Error (ppm)
Dilution
Single Dilution (C1 × V1 = C2 × V2)
Serial Dilution
| Level | Concentration |
|---|
Concentrations use whatever unit you enter. Keep stock and target in the same unit.