LC-MS Calculators

Method scaling, column volumes, adducts and mass accuracy, and dilutions. Results update as you type.

Scale a Method to a New Column

Keeps linear velocity and gradient column volumes the same. Same stationary phase assumed.

Current Method

New Column

Scaled Method

Pressure and plate numbers are relative estimates at the same solvent, temperature, and packing quality. Check your system's pressure limit before running.

Column Volume and Wash Volumes

Empty column volume = π × r² × L. Use it to turn "10 column volumes" into mL and minutes.

Column and Flow

Results

Quick Reference at This Flow Rate

Column volumesVolume (mL)Time (min)

What the Sources Say

So 10 column volumes is a sensible default for reversed-phase re-equilibration, at the upper end of Waters' 5 to 10, but it is not universal. Sources also count differently: Waters uses empty column volumes, while ACE's table of internal volumes works out to about 63% of the empty-tube volume. Confirm the number for your own method.

Adducts and Mass Error

Adducts from Neutral Monoisotopic Mass

Ionm/z

Uses ion masses (electron mass accounted for). Verify against your own software before reporting values.

Mass Error (ppm)

Dilution

Single Dilution (C1 × V1 = C2 × V2)

Serial Dilution

LevelConcentration

Concentrations use whatever unit you enter. Keep stock and target in the same unit.